Highlights

  • Versatile, wide range, multi-trip Temperature Logger
  • 16,000 readings memory capacity with up to 3 years battery life
  • OK’ & ‘ALERT’ Indicator with high and low customizable alarms.
  • Industry best download time – less than 5 seconds for full memory.
  • High performance at low cost

The LogTag TRIX-16 is a versatile, wide range, multi-trip Temperature Logger, featuring high resolution temperature readings over a measurement range of -40°C to +85°C (-40°F to +185°F). Enclosed in a robust and durable polycarbonate case, the TRIX-16 features a real time clock, which provides date/time stamps for each temperature reading.

Using the LogTag Interface and the freely available companion software LogTag Analyzer, the LogTag is easily set-up for recording conditions including delayed start, sampling interval, number of readings, continuous or fixed number of readings and configuration of conditions to activate the ALERT indicator. Readings are downloaded using LogTag Analyzer, which provides facilities for charting, zooming, listing data statistics and allows exporting the data to other applications such as Excel.

The LogTag TRIX-16 complies with the relevant international standards for temperature monitoring devices, such as FCC, CE, C-TICK, and RoHS. This not only demonstrates the quality of the LogTag® TRIX-16, but underlines its suitability for temperature monitoring applications where accuracy and consistency is required.

Product Model TRIX-16
Temperature Sensor Measurement range -40 °C to +85 °C (-40 °F to +185 °F)
Recorder Operating Temperature Range -40 °C to +85 °C (-40 °F to +185 °F)
Storage Temperature Range -30 °C to +55 °C (-22 °F to +131 °F)
Rated Temperature Reading Accuracy Better than ±0.5 °C for -20 °C to +40 °C.
Better than ±0.7 °C for -30 °C to -20 °C and +40 °C to +60 °C.
Better than ±0.8 °C for +60 °C to +85 °C.
Better than ±0.9 °C for -40 °C to -30 °C.
Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration
Rated Temperature Reading Resolution Less than 0.1 °C for -40 °C to +40 °C.
Less than 0.2 °C for +40 °C to +80 °C.
Less than 0.4 °C for +80 °C to +85 °C.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®
Sensor Reaction Time Typically less than 5 minutes (T90) in moving air (1m/s)
Recording Capacity 16,159 temperature readings. 112 days @ 10min logging, 168 days @ 15min logging
Sampling Interval Configurable from 30 seconds to several hours
Logging Start Options Push button start or specific date & time. Optional start delay of up to 18 hours
Recording Indication Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator
Alarms 1 configurable upper and 1 configurable lower alarms
Download Time Typically less than 10 seconds for full memory (16,159 readings), depending on computer or readout device used
Environmental IP65 (roughly equivalent to NEMA 4)
Power source CR2450 3V LiMnO2 Battery (Fixed)
Battery Life 1 year storage, followed by 2 – 3 years of normal use (based on 15 minute logging, download data monthly)
Real Time Clock Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
Connection Interface Interface Cradle
Software LogTag® Analyzer
Size 86 mm(H) x 54.5 mm(W) x 8.6 mm(T)
Weight 33 g
Case Material Polycarbonate
EN12830

EN 12830:2018 - Temperature recorders for the transport, storage and distribution of temperature sensitive goods - Test, performance, suitability

Wall Mount Bracket WMB-01
Wall Mount Bracket WMB-01

For all current LogTag® models excluding indicators and UTRED30/UTREL30 loggers

LTI-HID USB
LTI-HID USB

LogTag® LTI-HID Desktop USB Interface Cradle

Protective Enclosure
Protective Enclosure

The LogTag® Protective Enclosure is specifically designed for LogTag® recorder (...)