La LogTag® UTRIX-16 no requiere hardware especial ni software propietario para acceder a los datos registrados y puede generar un informe PDF totalmente detallado. Almacena lecturas de temperatura en tiempo real en un intervalo de medición de -30 °C a +70 °C, muestra alertas configurables por el usuario e incluye un conector USB integrado de larga duración. La UTRIX-16 está encerrada en la misma carcasa de policarbonato robusta y duradera común a otros productos LogTag.
Product Model |
UTRIX-16 |
Temperature Sensor Measurement range |
-30 °C to +70 °C (-22 °F to +158 °F) |
Recorder operating temperature range |
-30 °C to +70 °C (-22 °F to +158 °F) |
Storage Temperature Range |
0 °C to +40 °C (32 °F to +104 °F) |
Rated Temperature reading accuracy |
Better than ±0.5 °C (±0.9 °F) for -20 °C to +40 °C (-4 °F to +104 °F). Better than ±0.7 °C (±1.3 °F) for -30 °C to -20 °C (-22 °F to -4 °F) & +40 °C to +60 °C (+104 °F to +140 °F). Better than ±0.8 °C (±1.5 °F) for +60 °C to +70 °C (+140 °F to +158 °F). Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration |
Rated Temperature Reading Resolution |
< 0.1 °C/F LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag® |
Sensor Reaction Time |
Typically less than 7 minutes (T90) in moving air (1m/s) in EN12830:1999 |
Recording Capacity |
16,129 temperature readings. 56 days @ 5min logging, 168 days @ 15min logging |
Alarms |
1 configurable upper and 1 configurable lower alarms |
Download Time |
Typically with full memory (16,129 readings) in less than 30 seconds from time of insertion to availability of PDF report. Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured) |
Environmental |
IEC 60529: IP64 with USB cap fitted |
Power source |
CR2450 3V LiMnO2 Battery (Fixed) |
Battery Life |
1 year storage, followed by 2 – 3 years of normal use (based on 15 minute logging, download data monthly) |
Real Time Clock |
Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C) |
Connection Interface |
USB 2.0, A-type plug |
Software |
PDF Reader, LogTag® Analyzer 2.5 or higher |
Size |
93 mm(H) x 54.5 mm(W) x 8.6 mm(T) including protective USB cap |
Weight |
36 g |
Sampling Interval |
Configurable from 30 seconds to 18 hours |
Logging Start Options |
Push button start (with configurable start delay from 1 minute to 72 hours) or specific date & time |
Recording Indication |
Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator |
Case Material |
Polycarbonate |
Soporte de pared WMB-01
Para todos los modelos actuales de LogTag®, excluyendo indicadores y (...)
Caja protectora
La carcasa protectora LogTag® está diseñada específicamente para los productos (...)