LogTag TREL30-16CP

Standard Monitoring of articles stored at very low temperatures, where a remote probe is beneficial.

Standard Monitoring of articles stored at very low temperatures, where a remote probe is beneficial.

Highlights

  • Built-in audible alarm
  • Easy to read LCD display with 30 day summary
  • Up to 15,900 recordings - Enough memory for 66 days at 6 minute logging
  • Calibrated Smart Probe Ready
  • Record inspections at the push of a button

 


The LogTag® TREL30-16CP is the Smart Probe ready iteration of the LogTag® TREL30-16. When paired with the compatible LogTag® CP11 Smart Probe, you remove the need to relocate or return the LogTag® TREL30-16CP temperature recorder for calibration, ensuring continuous and uninterrupted logging.

The LogTag® TREL30-16 Temperature Logger measures and stores up to 15,900 temperature readings and has a measurement range from -90 °C to +40 °C (-130 °F to +104 °F).

The display is designed to show ‘at a glance’ if temperature excursions have occurred during the current day and up to the previous 29 days. The display also shows the current temperature reading, the current time, recording status and battery status.

The alarm is trigged if temperature readings are outside pre-set alarm limits. The status indicator changes from a ✔ to a ✖ and, if configured, the audible alarm will sound. The day marker will also appear if not already present.

The alarm status indicator and audible alarm can be easily cleared, but the day marker remains to ensure a visible record of excursions is shown on the display. A simplified alarm clearance, stop & start procedure is available as an option.

Details of any excursions in the preceding 30 days can be checked directly by inspecting the statistics history on the recorder’s display or in more detail by downloading the logged data via a standard LogTag® Interface cradle to LogTag® Analyzer operating software.

Please note: An interface is required for configuration and reporting

 

Product Model TREL30-16 (Replaceable Battery)
External Temperature Sensor Measurement Range -90 °C to +40 °C (-130 °F to +104 °F)
Recorder Operating Temperature Range -30 °C to +60 °C (-22 °F to +140 °F)
Storage Temperature Range -10 °C to +40 °C (14 °F to +104 °F)
Rated Temperature Reading Accuracy Better than ±0.5 °C for -20 °C to +10 °C (-4 °F to 50 °F).
Better than ±0.7 °C for -50 °C to -20 °C (-58 °F to -4 °F) and +10 °C to +30 °C (50 °F to 86 °F) .
Better than ±1 °C for -80 °C to -50 °C (-112 °F to -58 °F ) and +30 °C to +40 °C (86 °F to 104 °F).
Better than ±1.8 °C for -90 °C to -80 °C (-130 °F to -112 °F ).
Actual performance is typically much better than the rated values.
Accuracy figures can be improved by recalibration
Rated Temperature Reading Resolution 0.1 °C (0.2 °F) for measurements -80 °C to 0 °C (-112 °F to 32 °F).
0.2 °C (0.4 °F) for measurements 0 °C to 18 °C (32 °F to 64.4 °F) and -90 °C to -80 °C (-130 °F to -112 °F).
0.3 °C (0.5 °F) for measurements 18 °C to 30 °C (64.4 °F to 86 °F).
0.5 °C (0.9 °F) for measurements 30 °C to 40 °C (86 °F to 104 °F).
LogTag Analyzer® currently displays to one decimal place of °C or °F.
The native resolution is what is stored in the LogTag
Recording Capacity 15,905 temperature readings. 66 days @ 6min logging. Day summary statistics memory (for display on LCD): up for 30 days of Max/Min/duration and alarm trigger statistics
Sampling Interval Configurable from 30 seconds to hourly
Logging Start Options Push button start or specific date & time. Optional start delay of up to 18 hours
Recording Indication State indicator “REC”
Download Time Typically less than 10 seconds for full memory (15,905 readings), depending on computer or readout device used
Environmental IP61 (when vertically mounted or hung)
Power source 3V CR2032 (LiMnO2 ) coin cell battery, user replaceable
Battery Life Storage life of 1 year before ‘start’ (if it is hibernation mode), followed by: 1 year of operation with normal use (6 minute logging, statistics reviewed on the display no more than once daily for no longer than 30 seconds each time, download data monthly).
NOTE: Continual activation of the audible alarm will reduce the working life of the battery. When an alarm is triggered, the alarm should be cleared and reset as soon as possible
Real Time Clock Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
Remote Sensor Cable Type PTFE (FDA food contact rated) coaxial
Size 93 mm(H) x 54.5 mm(W) x 8.6 mm(T)
Weight 41 g without External Probe
Case Material Polycarbonate
Standard External Probe Cable Lengths Standard: 0.5 m (19”).
Extended: 3 m (9’10”) (recommended maximum)
Wall Mount Bracket WMB-01
Wall Mount Bracket WMB-01

For all current LogTag® models excluding indicators and UTRED30/UTREL30 loggers

LTI-HID USB
LTI-HID USB

LogTag® LTI-HID Desktop USB Interface Cradle

LTI-WM-WiFi Interface Cradle
LTI-WM-WiFi Interface Cradle

Upload data in real time via WiFi

Glycol Buffer
Glycol Buffer

The LogTag® Glycol Buffer is used where the temperature of (...)

External Probes and Smart Probes
External Probes and Smart Probes

All external probes and smart probes at a glance