LogTag® TRED30-16R Temperature Recorder with 30 day summary display and remote probe

Applications

Temperature measurement where access requires a remote probe and data needs to be accessed without a PC.

 

Product Summary

The LogTag® TRED30-16R Temperature Recorder measures and stores up to 15,900 temperature readings over a -40 °C to +99 °C measurement range, using a remote temperature probe.

The display is designed to show ‘at a glance’ if temperature excursions have occurred during the current day and up to the previous 29 days. The display also shows the current temperature reading, the current time, recording status and battery status.


The alarm is triggered if temperature readings are outside pre-set alarm limits. The status indicator changes from a ✔ to a X and, if configured, the audible alarm will sound. The day marker will also appear if not already present. The alarm status indicator and audible alarm can be easily cleared, but the day marker remains to ensure a visible record of excursions is shown on the display. A simplified alarm clearance, stop & start procedure is available as an option.


Details of any excursion in the preceding 30 days can be checked directly by inspecting the statistics history on the recorder’s display or in more detail by downloading the logged data via a standard LogTag® Interface cradle to LogTag® Analyzer operating software.

 

Product Highlights

  • Data logging Records time and temperature data for later download.
  • State Indicators Display shows current state (READY,STARTING,RECORDING, PAUSED or STOPPED).
  • Built-in Audible Alarm
  • Alarm Symbol Appears on the display to show an alarm trigger has occurred.
  • Day alarm summary Displays in calendar-like format alarm triggers for up to the last 30days.
  • Current time & temperature Shows the current time and the last recorded temperature when not in review mode.
  • Review button Allows review of current day and up to 29 previous days of statistical data and also adds an ‘inspection mark’ into the recordings.
  • Start/Clear/Stop button Push button recording start, alarm clear and recording stop (each option separately configurable).
  • Flexible start options Push button recording start with optional delayed start (up to 18 hours) or specific time & date start.
  • Flexible recording options Either record for a fixed time or record continuously with ‘wrap around logging’.
  • Rapid Download! Takes only seconds to download recordings.
  • Thin flat case Fits easily into packaging and thin enough to be easily mailed “letter rate”.
  • Real time clock Records time & temperature simultaneously.
  • 15.900 recordings capacity Enough memory for 66 days @ 6 min logging.
  • High Quality Connector Gold plated remote probe connector.
  • Interchangeable Probes LogTag® remote temperature ST100 probes are interchangeable.

 

Technical Specifications

Product model TRED30-16R (replaceable battery)
External temperature sensor measurement range -40 °C ~ +99 °C
Operating temperature range  -30 °C ~ +60 °C
Storage temperature range -10 °C ~ +40 °C
Rated temperature reading accuracy

Better than ±0.5 °C for -20 °C~+40 °C - typically ±0,3 °C
Better than ±0.8 °C for -40 °C~+20 °C - typically ±0,5 °C
Better than ±0.8 °C for +40 °C~+70 °C - typically ±0.7 °C
Better than ±1.2 °C for +70 °C~+99 °C - typically ±1 °C

Rated temperature reading resolution on LCD 0.1 °C for measurements -40 °C ~ +50 °C
0.2 °C(0.4 °F) for measurements +50 °C ~ +70 °C
0.3 °C(0.5 °F) for measurements +70 °C ~ +80 °C
0.4 °C-0.6 °C for measurements +80 °C~ +99 °C
Recording capacity Data logging memory : 15,905 logs - eg 66 days @ 6min logging, 80 days @ 15 min logging
Day summary statistics memory (for display on LCD): up for 30 days of Max/Min/duration and alarm trigger statistics
Sampling Interval configurable from 30 seconds to hourly
Logging start options Supports ‘continuous’ wrap around or specific recording period
Recording indication State indicator "REC"
Download time Typically less than 10 seconds for full memory (15,905 readings), depending on computer or readout device used.
Environmental IP 61 (when vertically mounted or hung)
Power source 3V CR2032 (LiMnO2) coin cell battery, user replaceable
Battery life Typically 1 year of operation.
NOTE: Typical battery life assumes 6 minute logging, day statistics are reviewed on the display no more than once daily for no longer than 30 seconds each time.
Download data monthly.
NOTE: Continual activation of the audible alarm will reduce the working life of the battery. When an alarm is triggered, the alarm should be cleared and reset as
soon as possible.
Real time clock Built-in real time clock.
Rated accuracy ±25 ppm @ 25 °C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006 ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
External probe cable lengths Standard: 1.5 m.
Extended: 3 m (recommended maximum).
Remote external probe type PTFE (FDA food contact rated) coaxial.
Size 93.0 mm(H)x54.5 mm(W)x8.6 mm(T)
Weight 43 grams without external probe
Case Material Polycarbonate
EMC Compliance Complies with EC EMC directives (EN 61000-6-3:2006 & EN 61000-6-1:2005)
Complies with FCC Part 15 Subparts A and B.
FDA 21CFR Part 11 Compliance Designed to comply with FDA 21CFR Part 11 (Digital signatures)
Reader interface  Standard LogTag® Interface cradle
Manufactured to meet the European RoHS Directive
LogTag® Recorders reserves the right to change this product’s specifications without notice.

 

Remote Probe Options

LogTag® TRED30-16R is designed and calibrated for use with the LogTag® ST100 remote temperature probe. Only this probe is compatible with TRED30-16R and using any other probe type is likely to cause an error in readings.

LogTag® ST100 sensors are designed and calibrated for interchangeability with TRED30-16R loggers to the standard rated accuracy of the product. If a TRED30-16R /ST100 probe combination is calibrated to a higher level of accuracy (as is possible) then care should be taken to always use the same TRED30-16R /ST100 combination for measurements as the ST100 sensor is not rated interchangeable at accuracies better than the TRED30-16R /ST100 rating in this specification.

.

Identifying an ST100 probe

The ST100 probe is distinguished from other LogTag® remote temperature probe types by the blue sleeve blue body at the plug end of the probe cable.

Stainless probe Ø 5 mm, short



ST100S-15 (1,5 m)
ST100S-30 (3,0 m)

The short tipped probe incorporates a 22 mm exposed length of 5 mm OD 316 stainless probe without handle but with small over mold to improve resistance to moisture and provide cable strain relief.
All materials at probe end of cable are FDA food rated including the cable.

Stainless probe Ø 3.2 mm, short


ST100J-15 (1.5 m)
ST100J-30 (3.0 m)

This short tipped probe incorporates ~40 mm of exposed length of 3.2 mm diameter 316 stainless probe.
All materials at probe end of cable are FDA food rated including the cable.
Supplied standard (part# ST100J-15) with 1.5 m (4'11”) or extended (part#ST100J-30) with 3 m (9'10") cable terminated to MCX plug.

Stainless probe Ø 5 mm, long


ST100T-15 (1,5 m)
ST100T-30 (3,0 m)

The tipped probe incorporates a 140 mm exposed length of 5 mm OD 316 stainless probe without handle but with small over mold to improve resistance to moisture and provide cable strain relief.
All materials at probe end of cable are FDA food rated including the cable.

Stainless probe Ø 3.2mm, medium


ST100K-15 (1.5 m)
ST100K-30 (3.0 m)

This medium tipped probe incorporates ~65 mm of exposed length of 3.2 mm OD 316 stainless probe.
All materials at probe end of cable are FDA food rated including the cable.
Supplied standard (part# ST100K-15) with 1.5 m (4'11”) or extended (part#ST100K-30) with 3 m (9'10") cable terminated to MCX plug.

Stainless handled probe Ø 5 mm


ST100H-15 (1,5 m)
ST100H-30 (3,0 m)

The standard handled probe incorporates a 70 mm long shaped overmolded handle with a 105 mm length 5 mm OD 316 stainless probe. All materials at probe end of cable are FDA food rated including the cable.

Stainless probe Ø 3.2 mm, long


ST100L-15 (1.5 m)
ST100L-30 (3.0 m)

This long tipped probe incorporates ~90mm of exposed length of 3.2 mm OD 316 stainless probe.
All materials at probe end of cable are FDA food rated including the cable.
Supplied standard (part# ST100L-15) with 1.5 m (4'11”) or extended (part#ST100L-30) with 3 m (9'10") cable terminated to MCX plug.

Bare probe


ST100B-15 (1,5 m)
ST100B-30 (3,0 m)

Bare sensor without stainless sheath ready for fitting to custom probe assembly. Not rated for food contact or environmentally rated.