LogTag UTRIX-16

Multi-use PDF Temperature Logger - satisfies the growing need for a cost effective and reliable multi-use temperature recorder

Multi-use PDF Temperature Logger - satisfies the growing need for a cost effective and reliable multi-use temperature recorder

Highlights

  • Automatically generates a PDF report and an Excel data file - no special download software required.
  • Automatically generates an encrypted LogTag data file - for a more detailed analysis and to comply with FDA CFR21 Part 11.
  • Fully user configurable via LogTag Analyzer for individual alarm settings, start options, trip length and log interval.
  • Durable, reliable USB connector with gold plated contacts, designed for multiple insertions.
  • Up to 16,129 recordings - provides over 3 months logging at 10 minute sample intervals.

The LogTag® UTRIX-16 requires no special hardware or proprietary software to access the recorded data and can generate a fully detailed PDF report. It stores real-time temperature readings over a measurement range of -30 °C to +70 °C, displays user-configurable alerts, and comes with an integrated, long-life USB connector. The UTRIX-16 is enclosed in the same robust and durable polycarbonate case common to other LogTag products.

Product Model UTRIX-16
Temperature Sensor Measurement range -30 °C to +70 °C (-22 °F to +158 °F)
Recorder operating temperature range -30 °C to +70 °C (-22 °F to +158 °F)
Storage Temperature Range 0 °C to +40 °C (32 °F to +104 °F)
Rated Temperature reading accuracy Better than ±0.5 °C (±0.9 °F) for -20 °C to +40 °C (-4 °F to +104 °F). Better than ±0.7 °C (±1.3 °F) for -30 °C to -20 °C (-22 °F to -4 °F) & +40 °C to +60 °C (+104 °F to +140 °F). Better than ±0.8 °C (±1.5 °F) for +60 °C to +70 °C (+140 °F to +158 °F). Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration
Rated Temperature Reading Resolution < 0.1 °C/F LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®
Sensor Reaction Time Typically less than 7 minutes (T90) in moving air (1m/s) in EN12830:1999
Recording Capacity 16,129 temperature readings. 56 days @ 5min logging, 168 days @ 15min logging
Alarms 1 configurable upper and 1 configurable lower alarms
Download Time Typically with full memory (16,129 readings) in less than 30 seconds from time of insertion to availability of PDF report. Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured)
Environmental IEC 60529: IP64 with USB cap fitted
Power source CR2450 3V LiMnO2 Battery (Fixed)
Battery Life 1 year storage, followed by 2 – 3 years of normal use (based on 15 minute logging, download data monthly)
Real Time Clock Built-in real time clock. Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
Connection Interface USB 2.0, A-type plug
Software PDF Reader, LogTag® Analyzer 2.5 or higher
Size 93 mm(H) x 54.5 mm(W) x 8.6 mm(T) including protective USB cap
Weight 36 g
Sampling Interval Configurable from 30 seconds to 18 hours
Logging Start Options Push button start (with configurable start delay from 1 minute to 72 hours) or specific date & time
Recording Indication Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator
Case Material Polycarbonate
Wall Mount Bracket WMB-01
Wall Mount Bracket WMB-01

For all current LogTag® models excluding indicators and UTRED30/UTREL30 loggers

Protective Enclosure
Protective Enclosure

The LogTag® Protective Enclosure is specifically designed for LogTag® recorder (...)