LogTag USRIC-16

Single-use PDF USB Temperature Logger

Single-use PDF USB Temperature Logger

Highlights

  • Integral USB connector - No separate Interface Cradle required
  • Automatically generates a PDF report - No special software required
  • Encrypted LTD file generated for a more detailed analysis
  • User configurable - No need to buy a special logger for each trip
  • Suitable for independent receivers, clinical trials & last-mile distribution

The LogTag® USRIC-16 PDF Temperature Recorder measures and stores up to 16,000 real time temperature readings over a measurement range of -30 °C to +60 °C.

Designed as a single-use, cost effective & reliable temperature recorder, the LogTag® USRIC-16, enclosed in a robust & durable polycarbonate case, automatically generates a fully detailed PDF report and provides the option to download the data using the freely available LogTag Analyzer software.

Using LogTag® Analyzer, the LogTag® USRIC-16 is easily set-up for a number of recording conditions.

Product Model USRIC-16
Temperature Sensor Measurement range -30 °C to +60 °C (-22 °F to + 140 °F)
Recorder operating temperature range -30 °C to +60 °C (-22 °F to +140 °F)
Storage Temperature Range 0 °C to +40 °C (32 °F to +104 °F)
Rated Temperature reading accuracy Better than ±0.5 °C for -5 °C to +30 °C Better than ±0.8 °C or better for measurements in other areas of rated range Actual performance is typically much better than the rated values. Accuracy figures can be improved by recalibration.
Rated Temperature Reading Resolution < 0.1 °C/F LogTag Analyzer currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®
Sensor Reaction Time Typically less than 7 minutes (T90) in moving air (1m/s) EN12830:1999
Recording Capacity 16,129 temperature readings. 56 days @ 5min logging, 168 days @ 15min logging
Alarms 6 user-configurable alarms
Download Time Typically in less than 20 seconds from time of insertion to availability of PDF report Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured)
Environmental IEC 60529: IP64 with USB cap fitted
Power source CR2032 3V LiMnO2 Battery (Fixed)
Battery Life 2 years storage, followed by 6 months normal use
Real Time Clock Built-in real time clock Typical accuracy ±25ppm @ 25°C (equivalent to 25 seconds/day) Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
Connection Interface USB 2.0, A-type plug
Software PDF Reader, LogTag® Analyzer 2.5 or higher
Size 93 mm(H) x 54.5 mm(W) x 8.6 mm(T) including protective USB cap
Weight 31 g
Sampling Interval Logging interval is configurable from 1 minute to 18 hours. (Typically 5 minute logging)
Logging Start Options Push button start or specific date & time. Optional start delay (1min to 72 hours)
Recording Indication Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator
Case Material Polycarbonate
EN12830

EN 12830:2018 - Temperature recorders for the transport, storage and distribution of temperature sensitive goods - Test, performance, suitability