LogTag TIC20

Tamper-proof WHO temperature indicator

Tamper-proof WHO temperature indicator

Highlights

  • Factory pre-configured to monitor a range of standard vaccines - 4 WHO PQS Pre-qualified models available
  • At-a-glance summary with ‘OK’/’Reject’ alarm status.
  • Review statistical data on device for up to 40 days prior.
  • Robust and Accurate. Data collected is secure and cannot be manipulated




Current temperature, elapsed time, load status, monitoring status and battery condition can all be seen at a glance at any time during the trip. On arrival at the destination, monitoring can be stopped and the user can then review time and temperature statistics for each day of the trip, up to 20 days maximum.

The LogTag TIC20 is factory pre-configured for one of a range of monitoring profiles. Special profiles are available on request, subject to conditions.

Statistical data can be downloaded from the TIC20 using an interface cradle and freely available LogTag Analyzer operating software if required.

The TIC20 offers an accurate and cost effective solution for monitoring the pharmaceutical cold chain in “last mile” distribution and other applications where full data logging capability is not required.

Product Model TIC20
Temperature Sensor Measurement range -30 °C to +60 °C (-22 °F to +140 °F)
Recorder operating temperature range -30 °C to +60 °C (-22 °F to +140 °F)
Storage Temperature Range -30 °C to +65 °C (-22 °F to +149 °F).
Rated Temperature reading accuracy Better than ±0.5 °C for -10 °C to +40 °C (+14 °F to +104 °F). Better than ±1.0°C for -30°C to -10°C (-22°F to +140°F). Better than ±1.0°C for +40°C to +60°C (+104°F to +140°F).. Actual performance is typically much better than the rated values.
Rated Temperature Reading Resolution Less than 0.1 °C for -30 °C to +60 °C.
Sensor Reaction Time Typically less than 7 minutes (T90) in moving air (1m/s)
Recording Capacity 1st activation of each alarm for each day including trigger time and duration of each excursion
Environmental IP64
Power source CR2032 3V LiMnO2 Battery (Fixed)
Battery Life Minimum storage life of 5 years before ‘start’. Monitoring period: 40 days. Minimum accessibility (display) period of 6 months after ‘stopped’
Real Time Clock Built-in real time clock. Typical accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
Size 73 mm(H) x 54.5 mm(W) x 8.6 mm(T)
Weight 30 g
Sampling Interval 1-15 minutes, factory set to 6 minute interval for profile 0006 (i.e. 0.1hour).
Logging Start Options Push button start or specific date
Recording Indication Alarms are displayed on a custom LCD
Case Material Polycarbonate

Ready-Mode Push any button. “Ready” shows the indicator is ready to be started.

Start Mode Press and hold the Start button. “Monitoring” will flash. Release Start button when flashing stops. The Indicator is now in the 60 minute start delay.

Monitoring Mode “Monitoring” appears on the display. The indicator is now taking a temperature reading every 6 minutes. Elapsed time in hours and days appears on the display. If an Alarm occurs, the day segment is illuminated and the OK symbol is replaced by an X.

Review Mode Press the Review button to view statistics for each day.

Stopped Mode Press and Hold the Stop button. “Stopped” will flash. Release Stop button when flashing stops. The Indicator is now Stopped.

Type 1 Alarm version DTP, DT, TT, Td, HepB, IPV, liquid Hib, MenAfrivac, HPV and PCV vaccines.
Type 2 Alarm version OPV, freeze-dried BCG, measles, MR, MMR, lyophylized Hib, yellow fever, meningitis, Rabies and Rotavirus vaccines.
Type Prevenar Alarm version Prevenar 7&13
Type RotaTeq Alarm version RotaTeq®